Note:
This project will be discontinued after December 13, 2021. [more]
Product:
Prismax_firmware
(Baxter)Repositories |
Unknown: This might be proprietary software. |
#Vulnerabilities | 3 |
Date | Id | Summary | Products | Score | Patch | Annotated |
---|---|---|---|---|---|---|
2020-06-29 | CVE-2020-12037 | Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device. | Prismaflex_firmware, Prismax_firmware | N/A | ||
2020-06-29 | CVE-2020-12036 | Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device. | Prismaflex_firmware, Prismax_firmware | N/A | ||
2020-06-29 | CVE-2020-12035 | Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration. | Prismaflex_firmware, Prismax_firmware | N/A |