Product:

M1000_multipara_patient_monitor_firmware

(Niscomed)
Repositories

Unknown:

This might be proprietary software.

#Vulnerabilities 4
Date Id Summary Products Score Patch Annotated
2020-08-26 CVE-2020-15484 An issue was discovered on Nescomed Multipara Monitor M1000 devices. The internal storage of the underlying Linux system stores data in cleartext, without integrity protection against tampering. M1000_multipara_patient_monitor_firmware 7.5
2020-08-26 CVE-2020-15482 An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the admin account. This allows an attacker to gain root access to the device over the local network. M1000_multipara_patient_monitor_firmware 7.8
2020-08-26 CVE-2020-15483 An issue was discovered on Nescomed Multipara Monitor M1000 devices. The physical UART debug port provides a shell, without requiring a password, with complete access. M1000_multipara_patient_monitor_firmware 6.8
2020-08-26 CVE-2020-15485 An issue was discovered on Nescomed Multipara Monitor M1000 devices. The onboard Flash memory stores data in cleartext, without integrity protection against tampering. M1000_multipara_patient_monitor_firmware 5.5